Laboratory Drying Oven – Forced-Air Hot Air Oven for Seed & Soil Labs
Forced-air laboratory drying oven for seed moisture-content determination, soil sample drying, and general lab heating. Digital temperature control up to 300 C with uniform heat throughout the chamber.
₨195,000.00
-
Pick up from the Woodmart Store
To pick up today
Free
-
Courier delivery
Our courier will deliver to the specified address
2-3 Days
Free
-
DHL Courier delivery
DHL courier will deliver to the specified address
2-3 Days
Free
-
Warranty 1 year
-
Free 30-Day returns
Payment Methods:
Description
The Laboratory Drying Oven is a forced-air hot-air oven that provides uniform, digitally controlled heating from ambient plus 10 C up to 300 C. A circulation fan moves heated air continuously through the chamber so every tray sees the same temperature – the critical requirement for reproducible drying and moisture testing.
For Pakistani seed and soil laboratories, the drying oven is the reference instrument for seed moisture-content determination (the low-temperature oven method), soil sample drying before analysis, glassware drying and heat sterilization, and general sample preparation. Seed moisture content decides storability, pricing, and rejection at intake; an oven with verified temperature uniformity lets the lab defend its moisture numbers to buyers and auditors.
Key benefits include PID digital temperature control with high accuracy, forced-air circulation for chamber uniformity, programmable timer and over-temperature protection, stainless-steel inner chamber that resists corrosion and is easy to clean, and multiple adjustable shelves for batch loading.
Typically supplied with the oven cabinet, 2-3 stainless-steel shelves, power cable, and operation manual. Temperature calibration certificate available on request.
Ideal for seed testing laboratories doing moisture-content analysis, soil testing labs, university chemistry and agriculture labs, food testing labs, pharmaceutical QC labs, and research institutes needing reliable drying and heating.
Key Features
- Forced-air circulation for uniform temperature throughout the chamber
- Digital PID temperature control, range ambient+10 C to 300 C
- Programmable timer and over-temperature protection for unattended runs
- Stainless-steel inner chamber – corrosion resistant, easy to clean
- Multiple adjustable shelves for batch sample loading
- Standard instrument for seed moisture-content (oven method) determination
- Also dries soil samples and sterilizes glassware by dry heat
- Runs on standard 220V / 50Hz supply
Specifications
| Capacity | 50 – 140 L (typical benchtop models) |
|---|---|
| Controller | Digital PID with timer and over-temperature alarm |
| Inner material | Stainless steel |
| Outer material | Cold-rolled steel, powder coated |
| Power consumption | ~1.5 – 2.5 kW |
| Power supply | AC 220V, 50Hz |
| Shelves | 2 – 3 adjustable shelves |
| Temperature accuracy | +/- 1 C typical |
| Temperature range | Ambient + 10 C to 300 C |
| Temperature uniformity | +/- 2.5 C or better at 100 C |
| Type | Forced-air convection laboratory drying oven |
How to Use
- Place the oven on a stable bench with clearance on all sides and connect to 220V power.
- Set the target temperature (e.g. 103 C or 130 C per the seed moisture protocol) and timer.
- Preheat the oven to set temperature before loading samples.
- Load pre-weighed samples on the shelves without blocking air circulation gaps.
- Close the door and run the full protocol time; the over-temperature alarm guards the run.
- Remove samples with heat-resistant gloves and transfer to a desiccator for cooling before re-weighing.
- After use, allow the oven to cool, wipe the chamber, and leave the door slightly ajar.
- Safety/maintenance: never dry flammable solvents, keep the fan intake clear, verify temperature with a reference thermometer quarterly, and check the door gasket yearly.